clear trend of deterioration in the figure of the platen. Normally, such
deterioriation might be attributed to the gradual release metallurgical
strains, but not in the case at hand, for the DBA platen was made of an
alloy called Precedent 71 which has a stability approaching that of fused
quartz. Some other physical process had to be involved. At this point,
the decision was made to measure the platen with film in place and with the
application of the film-flattening vacuum (about 0.1 atmosphere). This
vacuum was produced by the pump of the camera itself, for in this way there
would be no question concerning the correctness of the degree of the applied
vacuum. The results were startling. Relative to one corner of the format,
the center of the platen dropped over 20 um when the vacuum was turned on.
A full set of 169 measurements of spot elevations confirmed that a pro-
nounced, systematic change in the figure of the platen had indeed occurred
with the application of the film-flattening vacuum. Moreover, the actual
figure of the platen under the film-flattening vacuum could fully account
for the magnitude of the apparent radial distortion uncovered in the
Vermont and Atlanta Projects. Here, then, was the physical explanation
that had been sought. The deformation induced by the vacuum also explained
the slowly changing static figure of the platen — with continuing use the
platen had acquired an increasingly concave 'set' as result of repeated
flexing of the surface in response to each application of the vacuum.
Contour maps of the surface of the DBA Reseau Platen resulting
from the succession of measurements referred to above are shown in Figure
10. Of especial importance is Figure 10d which indicates the topography
of the platen when the normal operating vacuum is applied. It is something
close to this geometry that existed when the photographs of the Vermont and
Atlanta Projects were taken. One should appreciate, however, that the
laboratory measurements may not fully apply to the platen under operating
conditions in the camera, for to some extent the figure of the platen may
benefit from edge flattening resulting from pressure against the frame
defining the focal plane.
The experiments with the DBA Reseau Platen naturally raised
related questions concerning the operational stability of the Zeiss platen
normally used in the camera. Measurements before and after the application
of the vacuum were accordingly made on this platen also. The results given
in Figure 11 show that the Zeiss platen is subject to similar deformation
from the vacuum. The Zeiss platen seems also to have acquired a concave
set from long term use. As can be seen from comparison of Figures 10d and
11b, the contours of the Zeiss platen are more circular than those of the
DBA platen which tend to be rather square with rounded corners. These
differences reflect the respective geometries of the re-enforcing ribs on
the backsides of the two platens — a centrally radiating 'spiderweb'
pattern in the case of the Zeiss platen and a 3x4 'box cell’ pattern in
the case of the DBA platen. Because of the radial symmetry of the ribs
of the Zeiss platen, it turns out that a second degree polynomial can
provide an excellent analytical representation of the surface, whereas,
as can be seen from Table 3, a fourth degree polynomial is needed for a
comparable fit to the DBA platen. In the process of polynomial fitting,
it appears that terms involving even powers of x and y are dominant (note,
for instance, with the DBA platen the trivial improvement of the third
degree fit as opposed to the marked improvement with the fourth degree
fit). This result is in conformance with what is to be expected from the
series solution of the partial differential equation governing the defor-
mation of evenly loaded plates (Sechler (1968)). This consideration is
reflected in a revised version of the error model currently used at DBA,
a topic to be taken up later.
«D5.