(9
6.7 Aperture
Ihe apertures used in the measurements shall be stated in the
report.
6.8 Presentation of Results
The results shall be presented in the form of tables showing
the value of modulation transfer function (and phase transfer function
if measured) at each spatial frequency, for each field angle, azimuth,
and focal position tested. If preferred, results may also be shown
as graphs of transfer function versus spatial frequency for each
focal position, field angle, and azimuth. For some purposes, it
is best to use double logarithmic paper, but this is not universally
So, and the presentation of tabular data will allow the user to
select his own preferred method of plotting.
6.9 Guidelines for Measuring MTFs of Camera Systems
Standard procedures for measuring camera system performance
(camera-filter-film combination) in terms of modulation transfer
functions (MTF's should yield results which are representative
of camera system capabilities under both laboratory and operational
conditions. In conducting such evaluations, the following minimal
conditions should be considered:
6.9.1 Target: Edge (natural or man-made), bar, sinusoidal or
line targets may be employed, providing the contrast ratio between
the target and background is recorded on the straight-line portion
of the film D-log E curve. Contrast ratios of 3:1 to 6:1 are appro-
priate, 2 to 60 cycles/mm, in sufficient increments to provide
more than 5 data points. Edge targets should have a minimum image
size which is several times larger than either the effective slit of
the microdensitometer or the spread function of the photogrammetric
camera system. (Typical widths of photogrammetric camera system
Spread functions vary from 30 yum to 100 um depending on the film
employed.)
6.9.2 Sensitometry: Evaluations should be conducted using the
original film. Sensitometric data for the developed film should
be obtainable from density measurements of a step tablet impressed
on the film via a sensitometer prior to development.
6.9.3 Microdensitometry: Microdensitometer parameters should
be selected to avoid effects of coherence, noise, and instrumental
degradation of the system MTF. Typically, numerical apertures
of the condensing objective should be equal to, or larger than
that of the enlarging objective. Effective slit sizes of 1 um
x 80 um to 3 um x 500 um are recommended, with the smaller slits
reserved for use with fine-grain, high definition imagery.