Full text: Proceedings of the international symposium on remote sensing for observation and inventory of earth resources and the endangered environment (Volume 1)

  
    
  
  
  
  
  
   
  
   
  
   
   
  
  
   
      
  
   
   
  
-:591 - 
DESIGN OF A LARGE LAMBERT-TYPE REFLECTION STANDARD AND FULL-FRAME 
CALIBRATION OF A CAMERA LENS 
H.-P. BAHR 
Institut für Photogrammetrie und Ingenieurvermessungen, Technische 
Universitat Hannover 
ABSTRACT 
A 1.20 m by 1.20 m large BaS0, - reflector has been developed in order 
to provide a full-frame standard of perfect diffuse surface for light 
fall-off calibration of a 60 mm-HASSELBLAD BIOGON. The calibration was 
executed analytically by least-squares adjustment, using cosine func- 
tions with additional linear parameters, which lead to + 0.02 D rms error 
of the residuals. 
ZUSAMMENFASSUNG 
Zur flüchenhaften Bestimmung des Lichtabfalls in einem 60 mm-HASSELBLAD 
BIOGON wurde ein 1,20 m x 1,20 m grofer BaS0, - Eichstrahler mit voll- 
kommen diffus streuender Oberfläche entwickelt. Die Eichung des Objek- 
tivs erfolate durch eine Ausgleichung über Kosinus-Funktionen mit zu- 
sätzlichen linearen Anteilen, was auf mittlere Fehler von + 0,02 D 
führte. 
RESUME 
Un reflecteur - BaSO„ (1,20 m par 1,20 m) à surface parfaitement diffuse 
a ete devellopé pour calibrer la répartition de la lumiére à travers un 
objectif HASSELBLAD BIOGON (60 mm). La calibration fait intervenir 
l'ajustement de fonctions-cosinus par les moindres carrés, avec des 
paramétres linéaires supplémentaires. L'erreur moyenne sur les ecarts 
résiduels atteint + 0,02 D.
	        