the calibrated grid plate coordinates. This lookup table
of corrections is now used to produce calibrated stage
coordinates for any commanded stage position. This
calibration can be performed at any time and takes
about 30 minutes. Additionally, there is a "stage
verification" mode which measures a subset of the grid
intersections and tells the user if there is any problem
with the current calibration. This procedure takes about
5 minutes. The verification compares the calibrated
stage coordinates against the actual grid plate
coordinates and shows the user the RMS error in X and
Y. If this exceeds their desired threshold, the stage is
recalibrated. The stage verification procedure is
recommended about once a month and recalibration is
rarely necessary. Due to a well proven design, the
stage holds calibration for very long periods of time.
Sensor Geometric Calibration
The sensor image of the DSW 200 is also calibrated for
geometric accuracy to the same level as the stage.
This can be done by another glass grid plate that has a
grid line spacing of 5 mm and whose intersections are
also accurate to 1 micron. A 6 by 6 grid of intersections
will then appear on the focal plane of the 2k by 2k area
array when using an optical pixel size of 12.5 microns.
This grid data again permits us to automatically find and
measure each grid intersection to sub-pixel accuracy
using image processing. A geometric transformation
between the image pixel coordinate system and the
calibrated grid coordinate system in computed. This
transformation only allows uniform scale and rotation
between the sensor and grid. Deviations from this ideal
geometry are computed and output for display to the
user. No more than 3 microns rms is tolerated. This is
performed as a check on the setup of the camera to lens
to stage optical path. If there are any alignment errors in
the stages, lenses, or array, they are discovered during
this procedure. This verifies the camera and lens setup
for each new scanner.
Checking The Final Image
The scanned image is an array of pixels residing in the
computer or on the hard disk. The image can be
displayed and measured. lf we assume a maximum
scan of interest to cover about 220mm x 220 mm (9 inch
x 9 inch), then we should have a reference target of that
size to scan. A calibrated grid plate or film transparency
seems ideal, with intersections at a spacing sufficient to
uncover local distortion but not so frequent as to be
impractical to measure. The technology to produce such
accurate grids is readily available so no new wheel
needs inventing. This proposed reference would be
scanned and an image file produced. Next in this case,
every grid intersection would be measured by the human
placing a cursor on each perceived grid intersection or
by using image processing. When using the human and
a computer monitor, we would typically zoom the image
by a factor of 2 or 4 to help place the cursor to sub-pixel
positions on the grid intersections. However, an
algorithmic processing method would be preferred due
to its ease, speed and repeatability if one could be
agreed upon. If universally agreed upon algorithms were
employed, then a real scanner standard would begin to
be possible.
136
Let us assume the grid has intersections every 20
millimeters. That would require measuring about 100 grid
intersections over the entire image. With these
measurements, a 4 parameter transformation from
pixels to grid coordinates could be computed. This
transformation physically models rotation, translation,
and a scale change between the grid coordinate system
and the pixel coordinate system. The RMS of the
residuals from this transformation is a good indicator of
accuracy of the scanned image because the
redundancy is so high and testing was done throughout
the scanned image. This procedure can be performed
by us the manufacturer, or by the user. This permits a
statistically rigorous check on the resulting image
geometry directly.
RECOMMENDATIONS
Since the user of the image scanner or scanned image
data is really in control of the geometry of the image,
and since it is a relatively easy procedure to verify the
quality of a scanned image, we would recommend that
the user of the scanner test and produce a standard
accuracy statement. We would also recommend that
the manufacturer of the scanner provide the software
tools necessary to measure a calibration plate or film so
the user can produce the accuracy statement. We
would recommend that the professional associations
establish the specific methods by which this should be
done and the minimum content of the statement. The
manufacture or the user could supply the calibration film
or plate. A large 9 by 9 inch scan should be checked
with a minimum of 20 points spread throughout the scan
area. This would permit sufficient redundancy when
checking the quality using a 4 parameter
transformation. The resulting accuracy statement
should minimally state the RMS in X and Y directions
along with the maximum fit residual.
References
Leberl, Franz W., 1992, Precision Scanning Of Aerial
Photogaphy: ASPRS Technical Papers, 1992 ASPRS-
ACSM Annual Convention, Vol.1 pp. 247-252.
Mikhail, Dr. Edward M., 1992, Quality Of
Photogrammetric Products From Digitized Frame
Photography: 1992 ISPRS, International Archives Of
Photogrammetry And Remote Sensing, Vol. XXIX, Part
B2, Commission ll, pp. 390-396.
Sarjakoski, Tapani, 1992, Suitability Of The Sharp JX-
600 Desktop Scanner for The Digitization of Aerial Color
Photographs: 1992 ISPRS, International Archives Of
Photogrammetry And Remote Sensing, Vol. XXIX, Part
B2, Commission ll, pp. 79-86.
«
Presque
produit
plupart
Le bala
détermi
de tra
balayeu
l'étalc
soit ét
Mots-cl