562
Serie i:
eI Pu TN
(2) Calibration Lamp Data Analysis
agen rm
Internal calibration lamp data profile for each detector was obtained
by averaging data within every scan. By monitoring the data profile change,
an attemp was made to evaluate the calibration lamp data stability, linearity
and other characteristics.
ES
Monitoring was made for two scenes; (a) Detroit, July 25, 82, (b) Chicago,
Oct 25, 82. An example of the calibration lamp data profile (data (b), Band
4 Detector 8) is shown in Fig. 4, Features to be noticed are overshootings
which appear when the calibration lamp status transition occurs and periodic
level variation within each lamp status. Large overshooting (up to 20%) of
calibration lamp data was observed for Bands | to 4, while small overshooting
(up to 52) of calibration lamp data were seen in Bands 5 and 7. It is considered
due to the different detector types; i.e. Si photodiode for Bands | to 4 and
InSb photodiode for Bands 5 and 7. Periodic data level changes occur to all
the bands with periodicity of 6 scans. The exact cause of this periodic data
level change is not known.
Relationship between the averaged calibration value of each lamp status
and then lamp radiance is shown in Fig. 5. As shown in the figure, a good
linearity holds between the input radiance and then output calibration lamp
data. In the comparison of calibration data profiles between data (a) and
data (b), some decrease in gain was observed. The similar decrease in calibra-
tion lamp level was observed in all bands. Monitoring should be continued
to decide whether the decrease in gain is due to the detector responsivity
change or calibration lamp radiacne change.
ee SCAN-AVERAGED CAL DATA PROFILE «e so TH ABSOLUTE RADIOMETRIC CALIBRATION ss»
258 [ zur
rT I ROW DaF 540 16.« 3
BAND 3 OTCTR 8 DETECTR = 8
- A
288 228 A
| w
zi uv
g p
üi Au^ ui JY 25.82 Fa
-J 1sa = WS 2-31 À
= =i1sg, t
5 = “ uze
i g "^ YRS 23-3]
Ó 19g D 3 74
zm
= "f
7
sa |
si e
8
1 sa 198 158 298 258 388
SCAN NO, 8 à ‘ i
8 S 18 15 28
ee SCAN-AVERAGED CAL DATA PROFILE se LAMP RADIANCE ON/CM2-SR)
258 Fr
RECEPTION 821225
PATH 23 ROW 31
BAND 3 DTCTR 8 Left:Fig.4 Scan-averaged calibration
data profile for a scene.
upper: Detroit, July 25, 22
lower: Chicago, Oct. 25, 82
N
=n
0
158 |
Right:Fig. 5 Relationship between
lamp radiance and cali-
bration value.
(Band 3 Detector 8)
CAL DATA LEVEL
mJ
e
m
ss
1 58 188 158 288 258 3988
SCAN NO.