sults
r the
reci-
Was
cies.
]ved
sible
iccu-
5mm
data
low.
3ated
] that
n the
ld is
l(see
; Was
rtical
re are
arger
offset
eight
station film-based calibration networks were taken of
the test field; one before, and one immediately after the
video portion of the test.
igure
Re-furbished Test Field
A set of 120 video images was collected and measured
for this test. The 120 images were taken in two steps .
The first 60 images were collected by setting the
camera roll to 0° and by rotating the test field in 6°
increments on its rotating base. Similarly, the second set
of 60 images were collected with a camera roll of 180°.
Data analysis was performed using 8, 16,30,60and 120
station network sets of the collected data. Target im-
ages occupied approximately 100-150 pixels on the
Sensor.
Target images were driven to automatically using a
curve prediction algorithm. This required an operator to
initially identify targets on three frames by using a
pointing device and a graphics monitor. The remaining
frames were measured automatically at a rate of about
15 seconds per frame (including rotating the test field).
No prior knowledge of the object was required.
The results are divided into two sections. The first set of
results is from five different combinations of runs from
the set of 120 images measured. Analysis was per-
formed on 8, 16, 30, 60 & 120 station sets of this data.
The second set of results is a repeatability test composed
of portion of the 120 image set broken into eight
independent networks of 8 stations each.
Closures Results for Videk Camera (microns)
# of Stations | S s. Sad eomm
8 28 25 27 1/25
16 28 25 27 1/25
30 28 23 26 1/26
60 23 19 21 1/32
120 37 23 25 1/27
Table 5
The first set of results from this section is the RMS of
closures for the bundle adjustments for each of the run
combinations (Table 5).
Although these closures are consistent with previous
measurements of target images of approximately 150
pixels, pointing precision on the order of 1/50 pixel or
better for images of this size is expected. However the
thresholding technique suggested by (Trinder 1989) has
not yet been incorporated in the measurement process.
This may explain some systematic effects in the results.
For 256 grey levels Trinder suggests:
T=74.(SF)!*A-"!
where SF refers to the 2-sigma-width Gaussian
Spread function and A the target size
The second set of results is the estimated and actual
accuracy summary for each combination of runs (Table
6).
Estimated and Actual Accuracies for
Five Combinations of Runs (mm)
Estimated Actual
8 0.025 0.023
16 0.017 0.018
30 0.011 0.017
60 0.008 0.016
120 0.006 0.018
Table 6
As expected, the accuracy estimates improve propor-
tional to the square root of the number of stations. The
actual accuracies, however, show no significant im-
provement after 16 stations. This is not what would be
expected if the errors in the measurement were pre-
dominately random. It is therefore believed that
unmodeled systematic error remain in the data. Figure
5 shows a graph of both estimated and actual accuracies
asafunction ofthe number of stations. Also shownisthe
theoretical curve of accuracy improvement.
The graph does illustrate to a certain level that the
addition of multiple stations improves accuracy. In this
case accuracies improved from 1:50,000 in the 8
Station set to accuracies in excess of 1:70,000.
The second section of this test is a repeatability test
made by separating the 120 video images into eight
independent data sets. Each of the data sets was reduced