Full text: XVIIIth Congress (Part B1)

  
white square at the center of each cross. The squares were 
measured repeatedly at a Wild AC1 analytical plotter with an 
estimated accuracy of 2 - 3 um. 
  
  
  
  
  
Figure 1. Grid plates for geometric tests and calibration. 
4. TESTS AND CALIBRATION OF DTP SCANNERS 
Using the above test patterns the following five scanners were 
tested: Agfa Horizon, Agfa Arcus II, UMAX Mirage D-16L, 
UMAX PowerLook, Sharp JX-610. The first scanner belongs to 
our Institute and was tested over a few years, the remaining ones 
were tested at companies or were lent. In all tests a resolution of 
600 dpi was used (exception: 400 dpi for Mirage). The same test 
patterns, data measurement and analysis, and calibration 
procedures were used for all scanners. A difference exists for the 
grid plates. The 25.4 cm wide plates could not be put flat on the 
A4 scanners (Arcus II and PowerLook) and a part of the plates 
could not be imaged. Latter is important for the on-line plate 
because one of the border lines was totally missing. This plate 
positioning (one side of the plates was lying on the scanner frame 
around the scanner glass plate, less than 1 mm higher than the 
scanner glass plate) caused imaging displacements which could 
not be modelled by an affine transformation (as used in the 
interior orientation). For all geometric tests Least Squares 
Template Matching (LSTM) was used to measure the grid 
crosses. The standard deviation of the matched positions was 
0.03 - 0.04 pixels, i.e. 1.3 - 2.6 um, for the 600 dpi and the 400 
dpi scans respectively. In the following x-direction is the 
direction of the CCD line (horizontal), y-direction is the direction 
of the scanning movement (vertical). 
4.1. Geometric accuracy without calibration 
Table 2 shows the geometric accuracy of the scanners. For all 
scanners, except the Arcus, two scans were made. The results 
were similar for both scans, however here the worst of the two 
results is shown. For this test all grid lines were measured by 
LSTM and an affine transformation was computed between these 
values and the reference values (as measured at the analytical 
plotter). As control points either all points were used, or four 
corner or eight points. In the last two cases the remaining points 
were serving as check points and their errors are shown in Table 
2. The versions with all points as control show the global 
geometric accuracy of the scanners. Only for Horizon the 
accuracy is worse than 60 um, for the Mirage it is even close to 
20 um! The maximum errors are bounded and correspond to ca. 
2.5 - 3.5 RMS. The errors are generally larger in x, indicating 
large lens distortions. Using only 4 control points the errors of the 
check points increase. This is natural because the corner points 
have larger errors than points, let's say in the middle of the 
scanner stage, and thus the estimated affine parameters have 
larger errors. The big systematic errors introduced by the errors 
16 
of the corner points are also indicated by the large mean errors, 
which ideally should be zero. A version with 8 control points (4 
corners and 4 points at the middle of the borderlines) was also 
tested. The results were better, in some cases significantly. 
The above mentioned scanner accuracy may be sufficient for 
some applications. Consider for example a scanner with 100 
microns geometric error, used to generate hardcopies of digital 
orthoimages in scales 1:24,000 and 1:12,000, using 1:40,000 
scale input imagery scanned with 25 microns, and an orthoimage 
pixel size of 1 m (equal to the footprint of the scan pixel size). 
The scanner error translates to a planimetric error of 4 m in the 
digital orthoimage, and 0.17 mm and 0.34 mm in the 1:24,000 
and 1:12,000 hardcopies. This approximates the measuring 
accuracy in topographic maps, and may be acceptable for many 
users. 
Table 2. Statistical values (in um) of geometric accuracy without 
calibration. Errors (residuals) after an affine 
  
  
transformation. 
Control/ RMS Mean Max absolute 
Scanner | check 
points X y X y X y 
  
  
4/621 | 146 | 71 | -5 | -26 | 224 151 
  
Horizon | 8/617 | 147| 67 | -4 | -13 | 223 130 
  
625/0 | 92 | 54 | O 0 220 159 
  
4/621 | 106 | 51 | 67 | -39 | 214 117 
  
JX-610 | 8/617 91 | 42 | 45 | -26 | 182 105 
  
625/0 | 56 | 29 1 0 0 182 91 
  
4/621 35-120 | 24 |: 4 73 56 
  
Mirage | 8/617 32742091929" |*.7 67 54 
  
D-16L | 62510 | 18 | 19| 0 | © 56 51 
  
4/521 85 | 81 | S1 | -69 | 199 151 
  
Arcus 11 "8/5171 76% 1*62 1*36 "46 1“ 180 129 
  
525/0 | 63 | 41 | O 0 216 122 
  
4/546 | 101 | 112 | -66 | 103 | 181 177 
  
Power- | 8/542 | 87 | 77 | -45 | 65 158 138 
  
Look | 5500 | 52 | 43 | 0 | 0 | 185 | 114 
  
  
  
  
  
  
  
  
  
  
4.2. The geometric calibration procedure 
The calibration consisted of two stages. In the first stage the 
effects of the lens distortion were modelled. Radial lens 
distortion caused large displacements in x-direction, and the 
tangential lens distortion smaller but significant displacements in 
y-direction. The off-line plate was scanned, all points were 
measured by LSTM and an affine transformation between these 
measurements and the reference values using all points as control 
points was computed. The residuals of this transformation were 
indicating the occurring errors. These errors were transferred 
from the pixel to the scanner coordinate system. There an x- 
correction regular grid was interpolated based on the residuals. 
International Archives of Photogrammetry and Remote Sensing. Vol. XXXI, Part B1. Vienna 1996 
The sar 
grids w 
vibratit 
other sc 
availab 
distorti 
affine t 
(affine 
corres] 
subtrac 
left w 
subseq 
comput 
y-prec( 
stable, : 
(for the 
that we 
For the 
border 
affine t 
values 
residua 
errors ; 
border 
transfo 
swath » 
interior 
the im: 
borders 
are mai 
to the t 
the y-[ 
transfo 
We use 
in y). T 
can be 
used. 
4.3. Gi 
To che 
two pl: 
top of | 
This co 
that it i 
dense « 
plate w 
accomr 
not hap 
not lyir 
asymm 
scanne 
(unfort 
not pro 
line pla 
scanne 
positio; 
scanne 
terms ( 
scan of 
results 
Table ‘
	        
Waiting...

Note to user

Dear user,

In response to current developments in the web technology used by the Goobi viewer, the software no longer supports your browser.

Please use one of the following browsers to display this page correctly.

Thank you.