ed by the
^ with an
nology for
rior surface
parent that a
discussions
density and
nterpretable
ind corneal
about the
[he cameras
fitted with
itioned at
umeras were
type so that
his required
the limited
close to be
operational
iccess to the
lerate close
rcome these
beamsplitter
appeared to
ar technique
X measuring
The control
ugh a thin
n accurately
inated from
The control
he cameras,
1 onto the
ope. This
ight of the
] relative to
:
A variety of methods of marking the cornea were considered
including talc, as per Bonnet's work (Bonnet 1959, Le Grand
1961, Bonnet and Cochet 1962) and alternatives to talc,
particularly in combination with tear film stabilisers, as well
as the self-luminance approaches of El Hage (1972c),
Warnicki et al (1988) and Banda and Muller (1990). The
method adopted uses marked ultra-thin (10pm) hydrogel soft
contact lenses. This approach proved satisfactory for the
prototype although a teflon material used by Thall (1993,
1993 per com) would appear to have greater promise. The
indications are that it is simpler to use, it is likely to be
simpler to mark, or that a pattern of marks could be projected
onto it. The instrument uses two optic fibre bundles to
redirect a Mecablitz 45CT flash to the edge of the cornea
where much of the light is internally reflected behind the
cornea to back illuminate the contact lens. A third optic fibre
bundle from the same flash is directed to the back of the
control sphere.
Figure 2. The control points imaged off the beamsplitter
in the absence of an illuminated cornea.
The calibration and verification of the system was completed
using both the DLT and UNBASCI software (Abdel-Aziz
and Karara 1971, Karara and Abdel-Aziz 1974, Marzan and
Karara 1975, El-Hakim et al 1979). The DLT was used for
routine data reduction. The photocoordinates of the object
space control and the corneal reference marks were measured
monoscopically using a Zeiss comparator and correlated
prior to performing the camera calibration.
It was instructive to make some estimates of the expected
accuracy of the system and particularly to examine the likely
effects of varying parameters such as convergence angle,
base distance and object distance on the expected precision
of object space coordinates. The geometry of the system was
constrained by minimum object distances, by minimum depth
of field considerations, and by the obstructions caused by the
patient's brow and nose. The practical working range was
found to be an object distance of between 27cm and 39cm,
and convergence of between 15° and 30°.
Formulae presented by Abdel-Aziz (1974) and Karara and
Abdel-Aziz (1974) were used to investigated the expected
precision of the solution for a range of practical camera
geometries and the following conclusions drawn:
i. object space precision would be critically affected by
the image scale;
445
ii. optimum precision was likely to be achieved when the
camera axes were converging to a point behind the
cornea;
iii. for a given image scale, small changes in the camera
convergence would have little affect on precision; and
iv. for the best geometry, the object space precision in X, Y
and Z would be in the order of 15um, 151um and 20um
respectively, leading to an r.m.s. approaching 30um.
The design of the prototype allowed an experimental
evaluation of the accuracy and reliability of the system for a
selection of camera geometries. A glass sphere with a radius
of approximately 8.5mm was manufactured and its radius
measured to an accuracy of better than +0.5um using
standard optical interference techniques. The™ (X.Y)
coordinates of approximately 20 targets on the spherical
surface were measured and corresponding Z-coordinates
computed. Their accuracy was estimated to be better than
Sum. Because DLT was the algorithm used in routine data
reduction, it was used to photogrammetrically measure object
space coordinates for the test points. The measured
coordinates of the test points were then transformed to the
calibration coordinate system using a least squares three
dimensional rigid motion. Accuracy was assessed in terms of
the residuals on this transformation.
Figure 3. A cornea showing the target points marked
onto the surface of a 10pm thick contact lens.
Photogrammetric verification of the Keratocon indicated that
the accuracy achieved was consistent with expectations, with
the mean magnitude and r.m.s. of object space errors
approaching noise levels. Within the working range of the
instrument, predicted best precision in X and Y of
approximately 15um (s.d.) compared with measured accuracy
of between 3um and 10pum in X and 4um and 20pm in Y.
Predicted best precision in Z of approximately 20um (s.d.)
compared with measured accuracy of between 6um and18
um. The predictions based on the formulae of Abdel-Aziz
(1974) were therefore a useful indicator of accuracy.
Clinical verification of the Keratocon proved to be difficult,
primarily because there is not a more accurate method of
measuring corneal topography and because of the difficulty
of testing repeatability. There are two aspects of the
instrument's clinical accuracy that still have to be quantified:
i. It is difficult to quantify the accuracy with which the
contact lens is representing corneal topography. A
International Archives of Photogrammetry and Remote Sensing. Vol. XXXI, Part B5. Vienna 1996