Table 9• An example of the values of the adjustment in an A7 autograph (z - 300 mm).
24
Point or
Readings
Corrections
Settings to
Action
Control-
element
the scales
readings
x 5
500.200 mm
dt = 50 pm
500.300 mm
moving the floating mark
500.300 ram
y 5
50C.300 «
dl = 50 ”
500.200 "
over point 5 with the
adjusting screws of the
objective 0 p .
500.200 "
*5
500.300 mm
dXp=-100 pm
500.100 mm
moving the picture carrier
500.100 ram
500.200 n
dy p = 25 "
500.250 w
support with the adjusting
screws ^ until the floating
500.250 "
mark reaches point 5.
c
150.000 mm
dc = -100pm
149.900 mm
loosening the grub screws,
turning the principal
distance drum until 150.000
ram is read( without the
spindle turning) and
tightening the grub screws.
150.000 ran
¥
loofboo
0
0
R
LTV
II
100 §35 5
loosening the grub screws,
loofboo
CO
100 §000
=-400 GG
99§960
setting to 100§000 (without
the spindle turning) and
tightening the grub screws.
100§000
bx
0.000 mm
dbxs=-100 pm
base out
loosening the grub screws,
0.000 mm
0.100 ram
setting to 0.000 ram
by
0.000 11
dby*= 200 n
11
H
&
(without the spindle
0.000 «
0.200 ”
turning) and tightening
bz
0.000 ”
dbz= 100 «
bZi =
99.900 "
the grub screws.
0.000 "
see Instructions for Wild A7-