e a . + , , WAVELENGTH (NN)
vos sm. o 78. d
DRTE: 788783 TIME: 12.27 - 12.28 RRER: 15
Fig. 3. Calculated reflectance from data in fig. 2. * | standard
deviation is plotted at each measurement point.
method used is the same as mentioned above. The instrument factor k for
each wavelength interval is derived from
v Od r
OR NE
where p is the bidirectional reflectance, Jr the directional reflectance
from the known reference panel, L, the instrument value of the radiance
from the panel and E; the instrument value of the irradiance on the diffus-
ing disc.
Rejection of spurious instrument values
In a set of measurements some of the values can be anomalous, as seen in
fig. 2. If these outlying values are wrong and accepted or right and rejected,
the computed reflectance value will be biased. Hence, it is desirable to find
an objective criterium for acceptance or rejection of values. Several sta-
tistical methods have been proposed by e.g. Anscombe (1960), Bliss et al
(1956) and Grubbs (1950). However, very few values are sampled at each wave-
length interval during a set of radiance measurements. This makes it
questionable to use probability tests. As the energy distribution of the
spectrum is relatively constant but the effect varies, the ratio have been