1024(V) pixels with a pixel size of 16 micron
and a focal length of about 20 mm.
The calibration of the stereo vision system was
done using a targeted test field. The stereo pairs
were taken at different distances and view
angles. Thirteen control points and four check
points were used.
3.1.1 Calibration with and without the
Relative Orientation Constraints
This test was conducted to analyze the relative
orientation constraints. Eight image pairs were
used for this purpose. The adjustment was
calculated twice; with and without the relative
orientation constraints. The relative orientation
parameters were derived for every image pair
from the exterior orientation parameters
obtained by the bundle solution. Tab. 1 and Tab.
2 show the calibration results with and without
the relative orientation constraints. Without the
constraints, the relative orientation parameters
are different from one image pair to another and
they are inaccurate.
The base length between the stereo cameras
obtained with the relative orientation constraints
is exactly the same as the length measured by
tape ( 1.827m + 1 mm). It is obvious that the
calibration with the relative orientation
constraints gives the stable parameters. The
relative orientation constraints improved the
results especially when the geometry of the
blocks and the distribution of control points are
not good.
3.1.2 Calibration with and without
Additional Parameters
In this case, the calibration of the stereo vision
system was done using the eight digital stereo
pairs of our test field. The bundle triangulation
with the relative orientation constraints was
computed twice, with and without additional
camera parameters. The principal point and the
focal length of both cameras were always
treated as unknowns. Seventeen points are used
in the adjustment, where thirteen were control
points and four were used as check points to
compare the accuracy. Results are presented in
Table 3.
The additional parameters improved the
accuracy by a factor of two. The radial
distortion parameter k1 is the most significant
parameter.
3.1.3 Position Accuracy of the Stereo
Vision System
We used the computed orientation parameters in
an intersection program to determine the object
coordinates from the image coordinate
measurements. This corresponds to the
positioning of points with the stereo-vision
system of the GPSVan, independent of object
space control. Again, the coordinate of the
targets of the test field were used for
comparison. The results are displayed in Tab. 4
showing the RMS errors positioned from
different distances.
It is fair to say that the positioning accuracy
with two Kodak DCS cameras are in the 10 cm
level for objects closer than 30 m in front of the
van.
3.2 Global Positioning Accuracy
To test overall accuracy of our mobile mapping
system, we measured four control points with
a global accuracy of *1cm. To check the
positioning accuracy, all control points were
located by the mobile mapping system. Table 5
shows the difference from points positioned by
the GPSVan's stereo vision system and their
true values. The distances of points to the
GPSVan are shown. All points are in the State
Plane coordinate system (Ohio south, zone code
3402).
For this test, the Turbo Rogues GPS receivers
were installed on the GPSVan. The rotation
angles of the GPSVan were taken from the
combined adjustment of GPS and inertial system.
The offset between the stereo vision system and
positioning systems was calibrated using the
method discussed in section[2.3].
Image Pair Bx [m] By [m] Bz [m]
AQ Ao AK
All 1.826 -0.008 -0.060
-O:
04297 0.05933 0.02015
Tab. 1 Relative orientation parameters obtained by adding the
relative orientation constraints in a bundle solution
484
| Image |
200
201
202
203
204
206
208
209
In this p
algorithmi
positionin«
one foot i
the req
applicatio