s to write the
derived from
(7)
om system of
Ww; (8)
W,
(9)
X», Yo, Zo and
as a test field,
m x 1.8 m. The
ets density in
1 with original
ide robust and
n in the image
er.
; points allows
rior parameters
the procedure is
ra in required
The image of the test field with recognized coded targets is
shown in Figure 3.
Figure 3. Calibration test field
Original calibration procedure (Stepanyants D., 2000) for low
precision test field is used. For calibration procedure 16 images
of test field are captured (8 images for each camera). The test
field is located in front of camera approximately at average
distance between camera and test field about 8 m thus providing
image capture at infinity adjustment. Images are captured for
angle positions of cameras of a=0,+45°, wx0,+45°, k=0, 180°.
Some of the images at different angle positions acquired
for calibration are shown in Figure 4.
Figure 4. A samples of test field images for calibration
Image coordinates of reference points marked by coded targets
are determined automatically with sub-pixel accuracy by ellipse
operator. The additional parameters describing CCD camera
model in co-linearity conditions are taken in form:
Ax - Ax, Xr Ky e Xr Ke Xr9Ks (2 2x7) 2xyP, m
Ay =Ay, + yr Ky 5r* Ky + rOKy + 258 + (2 + 2y2)P,
EM AX, iy ==-m (y—¥, )ir= |x’ + 5°
where x, y, — the coordinates of principal point,
m,, m, — scales in x and y directions,
K,K,K; — the coefficients of radial symmetric
distortion
P,P, — the coefficients of decentering distortion.
These parameters of interior orientation are to be estimated.
The available means for producing test fields do not allow to
provide desired accuracy of independent determination of
reference point coordinates. Only the relative distances between
reference point could be measured with required precision of
2.0 mm. Therefore plane test field is used for calibration.
Table 5 presents distances between given reference points used
for calibration and residual errors in reference distances after
bundle adjustment procedure.
# Reference Distance, Residuals,
point # mm mm
0 1-36 1644 -0.1257
1 6-30 1040.5 -0.0017
2 1-6 1000.5 0.0003
3 6-36 1300.5 0.0020
4 1-30 1447.5 -0.0000
5 30-36 260 -0.0015
6 8-29 986.5 -0.1159
7 11-26 986.5 0.0096
8 8-11 603 -0.0058
9 11-29 781 -0.0079
10 8-26 781 -0.0072
11 26-29 602.5 -0.0056
12 1-29 1315.5 0.1256
13 8-36 1315.5 0.1252
14 11-30 806 0.0002
15 6-26 1315 -0.0004
Table 5. Reference distances used for calibration and residual
errors in reference distances after bundle adjustment
The estimated parameters vector includes:
e exterior orientation parameters of all the images
e interior orientation parameters of both cameras
* coordinates X, Y of all the object reference points
excluding two points determining object space
coordinate system. With the plane test field
assumption Z coordinates of all points is taken to be
Zero.
The first stage of calibration procedure (image set acquisition)
is not automated now and the test field is installed at given
position manually. After that calibration procedure is executed
automatically performing the following steps:
e reference points recognition and their image
coordinate determination in all acquired images;
* initial values for exterior orientation parameters
definition;
51