Figure 1 “time 2" time 3" time 4^ time 5^ time Simulation
data data data data data data
Depth Accuracy Fig.10 | 0.33mm 0.41mm 0.46mm 0.39mm 0.48mm 0.74mm
(Difference of K, from the 1* time) (0.0) (8.10e-5) (1.18e-4) (4.25e-5) (1.36e-4) (3.0e-4)
Depth Accuracy Fig.11 0.38mm 0.52mm 0.46mm 0.39mm 0.41mm 0.68mm
(Difference of K; from the ]* time) (0.0) (-9.79e-6) | (4.74e-6) (1.29e-6) | (-2.23e-6) (2.0e-5)
Depth Accuracy Fig.12 | 0.38mm 0.41mm 0.41 mm 0.38mm 0.38mm 0.74mm
(Difference of x, from the 1* time) (0.0) (12um) (-14um) (-7um) (-8um) (70um)
Depth Accuracy Fig.13 0.38mm 0.38mm 0.43mm 0.41mm 0.41mm 0.79mm
(Difference of y, from the 1* time) (0.0) (-6um) (21um) (-15um) (17um) (70um)
Table3. The non-uniformity of accuracy by the reproduction
And the Table 3 shows the accuracy of this reproducibility. In
order to assess the accuracy of the Flat Sheet calibration we
made 3D measurement of the control points of 3D Target-field
by using the calibration parameters obtained by the Flat Sheet
DC1000-processed calibration (f — 7.2mm, Camera length:
1313mm, Base length: 528mm), where the accuracy is the
lowest. As you notice, everything beautifully falls within the
range of 0.5mm (about 1/3000).
The data in the far right column is the result of the simulation
case where the non-uniformity of data got worse more than two
fold (Simulation data 7.2mm in the graph). But even in this
simulation the accuracy was still as high as 0.7mm.
4. 3D MEASUREMENT BY ON-SITE CALIBRATION
Based on the results so far obtained, we went to the actual site
of a stone piling wall and performed 3D measurement (Figure
14). We tested at 3 points within the focal length range
between 72.mm and 50mm: one at comparatively wide angle
(9mm), the second near the middle (30mm) and the third at
comparatively telephoto angle (42mm). The size of the
photographing area was about 2mx2m.
The photographing conditions are on the Table 4.
Figurel4. Stone piling wall (Test field)
The precision actually obtained is in the Table 5.
We succeeded to obtain the satisfactory precision which is
within 2mm for the targeted lcm. We can also note that the
>>. Focal Camera Base accuracy was much higher (in our experiment nearly 10 times
length | distance H | length B higher) than when we did not make calibration.
Casel: wide (mm) 9 2586 774
Case2: middle (mm) 30 6494 1904
Case3: telephoto (mm) 42 7516 1706 oxy = H/f*op (3)
ôz = H/f*H/B*6p (4)
Table4. Conditioning for photographing and analyzing
We made the test following the flow chart of Figure 12, using where &xy = resolution of one pixel on surface
PI-2000 for 3D measurement. We also measured 49 points with &z = resolution of one pixel in depth
our Total-Station (measuring accuracy: 4—1mm), to 8 points of H = camera distance, B = base length
which we applied Self-calibrating bundle adjustment to f = focal length, Jp = pixel size
calculate the focal length, while the rest of the points served to
assess the accuracy.
Table5. The result of accuracy test
Focal Position Focal length (Analyzed | y-Parallax at relative Surface accuracy | Depth accuracy Óxy oz
value) [mm] orientation [um] [mm] [mm] [mm] [mm]
Casel: wide 8.83 1.5 0.7 1.4 1.0 3.3
(with no correction) (8.8) (8.4) (13,5)
Case2: middle 30.04 0.8 1.3 1.8 0,7 2.5
(with no correction) (6.7) (9.6) (15.2)
Case3: telephoto 43.21 1.3 1.4 1.9 0.6 2.6
(with no correction) (6.8) (11.3) (18.8)
(calibration data in the widest angle : 7.2mm)
58.