International Archives of the Photogrammetry, Remote Sensing and Spatial Information Sciences, Vol XXXV, Part B7. Istanbul 2004
2. MEASUREMENTS
A spectral measurement always follows the same scheme:
l. Measurement of light source
2. Measurement of cuvette
3. Measurement of light source
This sequence guarantees for the stability of the light source
and the sensor. Measurements with a duration of several hours
have shown a constant behavior of the system.
2.1 Results of the reference measurement
The first measurement of the holmium oxide cuvette showed
the following results: (The figure of the spectral optical
measurement is shown in Appendix A)
EROR COMPRO Measured peaks Differences
279 nm Not measurable Not measurable
361 nm 362 nm - I nm
453 nm 450 nm +3 nm
536 nm 537 nm - 1 nm
638 nm 639 nm - 1 nm
Table 3: Absorption peaks of the holmium oxide cuvette
The comparison between a priori known values and measured
absorption peaks shows maximum deviations of approximately
3 nm. This corresponds with the manufacturers information of
the spectral resolution of the sensor so that the calibration
which transforms sensor position into wavelengths can be
accepted.
Measured intensities have then been tested by using the neutral
glass filters, tab. 4 and Appendix B.
but a systematic offset of —0,014 is statistically significant.
Nevertheless, considering this offset does not improve the
derived spectral optical density data to a remarkable amount.
2.2 Results of the repetition of the measurements
During field work the instrument is subject to environmental
strain of climate like temperature and humidity, vibration and
others. So, after four months of use in laboratory and field work
a review of the calibration of the sensor has been undertaken. In
order to minimize error sources in the repetition the same fiber
optics have been used in the same sequence as in the first
measurement.
Tueereed peaks Measured peaks Differences
279 nm Not measurable Not measurable
361 nm 359 nm + 2 nm
453 nm 448 nm + 5 nm
536 nm 537 nm - 1 nm
638 nm 639 nm - 1 nm
Table 5: Results of the holmium oxide cuvette after 4 months
The results did not show significant differences to the previous
measurements. The narrow banded absorption peaks of
holmium oxide have been found again with differences in the
order of magnitude of the spectral resolution of the sensor.
Also, the systematic offset of the ordinate has been repeatedly
determined with —0,016.
Measurements
Wavelength n(F2) n(F3) n(F4)
440 0,288 0,516 0,987
465 0,260 0,476 0,928
546 0,256 0,482 0,942
590 0,269 0,508 0,977
635 0,269 0,495 0,934
Differences
Wavelength n(F2) n(F3) n(F4)
440 -0,020 -0,017 -0,023
465 -0,023 -0,000 -0,029
546 -0,018 -0,013 -0,016
590 -0,014 -0,008 -0,017
635 -0,014 -0,009 -0,019
Measurements
Wavelength n(F2) n(F3) n(F4)
440 0,282 0,514 0,987
465 0,255 0,474 0,923
546 0,254 0,480 0,937
590 0,268 0,506 0,072
635 0,268 0,492 0,929
Differences
Wavelength n(F2) n(F3) n(F4)
440 -0,014 -0,015 -0,023
465 -0,018 -0,017 -0,024
546 -0,016 -0,011 -0,011
590 -0,013 -0,006 -0,012
635 -0,013 -0,006 -0,014
Table 4: Results of the neutral glass filters
The results are convincing in the whole visible spectral range
and they are in accordance with the manufacturers information
90
Table 6: Results of the neutral glass filters after 4 months
2.3 Measurement of chlorophyll
The reason for considering the concentration of chlorophyll in
plants in a Phytoremediation project results from the influence
of pollutants, e.g. heavy metals like lead or cadmium, on the
production of specific kinds of chlorophyll in plants. Thus,
chlorophyll can serve as an indicator for the occurrence of
pollutants in soil and is expected to allow for the estimation of
Internation
ME À
quantities (
plants.
Chlorophyl
use of ligh
with which
and a col
(http://ww^
There are
absorption
e Chl a:
e Chb:
e (uU Chl c:
e Bacte
In photos:
electron in
molecules,
collecting :
In land pla
techniques
molecules
and under
triturated i
leaf struc
acetone. T
clear, dee;
decanted ti
The chlor
measured
from (Arn
CHL(a) -
CHL(b) -
For contr
TriOS ser
Elmer U
UV Winlat
above the
The result
% of the €
system an
Furthermc
0,014 in t
significan:
When acc
parameter
illuminati:
measurem
sensor sys
By using
resolution
easily ver
optic cabl
avoid any