cameras allows the storage of a large number of images
without the need to download them (up to 48 in the Apple
QuickTake). Although this is a necessary and useful
function, the compression techniques modify the image.
The modifications are not noticeable to the human eye,
but can affect the results from photogrammetric
measurements. Digital still cameras usually provide the
option of low resolution images (high compression) and
high resolution images (low compression). All images
taken for calibration purposes have been acquired as
high resolution images.
4. PROCEDURE
The calibration procedure developed would best be
described as a self-calibrating test range calibration, as
opposed to laboratory or on-the-job calibration (as
described by Osborn, 1994). A target plate was
manufactured (see Figure 1), being very similar to target
plates used recently in other calibration procedures (for
example: Clarke et al, 1995; van der Vlugt, 1995).
However, instead of using relatively small targets (2mm
dia) which would produce images of targets of 5-7 pixels
diameter, the targets were larger (5mm dia) to produce
images of up to 30 pixels diameter, depending on the
object distance used (ranging from 0.8m to 1.5m). This
approach follows that of Shortis et al., (1995b), as it
provides more reliable results for the centroiding of the
targets.
NI
E d
w
®
æ
Figure 1. Target Plate.
The targets were punched from a sheet of retro-reflective
material following research by Zumbrunn (1995), who
showed that punching out the material was the preferred
method of target manufacture. If a target is made by
masking the retro-reflective material, the image of the
target would be systematically shifted when viewed at a
large angle (over 30°), thus making targets produced by
that method unsuitable for this procedure.
The target plate was spray painted matt black to provide
contrast to the retro-reflective targets. The effectiveness
of this can be seen in Table 2 in Section 5 of this paper,
which shows the background in the images (average
minimum grey value) to be very low and the image of the
targets (average maximum grey value) to be extremely
high. The grey values (GV) can range from O to 255.
The target plate was coordinated using 1” theodolites to
an accuracy determined to be 0.3mm. This was
140
confirmed during the bundle adjustment and also by the
photogrammetric tests of a fellow researcher (Mustaffar,
1995).
An extension to this target plate was constructed, to
provide flexibility to accommodate the different focal
lengths of the cameras. The original target plate as
shown in Figure 1, fitted in the centre of the extension
plate, which measured 700mm by 500mm. The
calibration of the Philips cameras only required the
original target plate to be imaged, whereas the Logitech
Fotoman and the Apple QuickTake required the larger
target plate to be used, due to their shorter focal lengths.
Images of the target plate were acquired using each of
the four cameras to be calibrated. The images were
acquired from nine different camera stations positioned
around the target plate at 45? intervals and also from
vertically above. The approach taken follows the self-
calibration procedure described by Shortis et al., (1995a).
For each of the two analog video cameras, the camera
was set up on a tripod, warmed up for a sufficient period
of time and then a set of ten images was taken at each of
the nine camera stations. These images were averaged
to provide a final set of images to be measured for use in
the bundle adjustment. As a check, one set of the
original images of the Right Philips was also measured to
determine the effect of the averaging and is referred to in
the results as Single Frame.
PC-based software programs were written to average the
images, to make radiometric corrections to the images
and to measure the image coordinates of the targets. A
literature survey showed that there is usually no
radiometric correction done to images after they have
been acquired for errors such as pixels in the CCD array
which show black when they should show white or the
reverse.
As an adjunct. to the calibration procedure, an
investigation was carried out to determine whether
corrections for such ‘bad’ pixels should be incorporated
into the procedure. Such pixels could include isolated
ones, blocks of pixels or the array as a whole not
responding uniformly. This has been referred to as Non-
Uniform Pixel Response by Beyer (1992). The result of
the investigation is given in Section 5 of this paper.
A PC-based computer program (see Fryer, 1994 for a
description) was used for the measurement of targets.
The program was menu driven and displayed the digital
image to be measured. To measure a target, the
operator defined a rectangular area around the target,
ensuring no other target encroached on that area. This
area was processed to find the centroid of the target.
The centroiding involved accessing each grey value (GV)
in the given window, finding the average GV and the
maximum GV and establishing a threshold level. All
pixels with a GV below the threshold level were ignored.
The average threshold level for each target is shown for
each camera in Table 2 in Section 5, along with the
average number of pixels falling above the threshold level
for each target.
International Archives of Photogrammetry and Remote Sensing. Vol. XXXI, Part B1. Vienna 1996
Thre
usec
sque
one
pixel
valut
grey
valu
effec
mett
each
resu
Sect
grap
met!
The
files
bein
usec
ever
was
Prioi
adju:
P2
Thes
adju:
the
men
The
tabu
calib
the |
and
for ‘I
cam
The
Tabl
imac
offse
para
free(
it ca
cam
RMS
how:
the i
foca
cam
relat
The
5-1C
plun
The
Sect
thes
anal
sign