Full text: XVIIIth Congress (Part B5)

  
be generally the case and is certainly not the case for a 
keratoconic cornea or irregular astigmatic cornea. 
e Keratometer measurements become increasingly difficult 
when surface irregularities distort the mires. 
e The point-by-point nature of the measurement process 
makes it impractical to compile a complete mapping of 
corneal topography. 
The keratometer fulfils its primary role, that of providing a 
radius of curvature or, on the basis of an assumed corneal 
refractive index, a value of the corneal refractive power, for 
routine clinical assessment of normal human corneas. Its 
application beyond this role is very limited. 
Photokeratoscopes are also, in many respects, inadequate: 
* The central and peripheral regions of the cornea cannot 
normally be measured because of an absence of reflected 
mires, corneal curvature, and obstruction caused by the 
nose and brow (Bores 1991, Gormley et al 1988, Klyce 
and Wilson 1989a, Mammone et al 1990, Warnicki et al 
1988). Accurate measurement of the central reflected mire 
is critical to the mathematical derivation of corneal shape 
(Missotten 1994). 
e The location of reflected mires is highly sensitive to 
corneal distortion. This is because the position of the 
reflected mire will be a function of both slope and 
displacement. On normal corneas, this sensitivity can be 
advantageous. | However when there are very rapid 
changes in topography the image soon becomes too 
distorted to measure reliably (Klyce and Wilson 1989b, 
Friedlander et al 1991). 
e Visual interpretation of a keratogram can only provide 
approximate data; clinically significant changes are 
commonly not detectable (Friedlander et a/ 1991, Wilson 
and Klyce 1991). 
e Several commercial instruments use dithering techniques 
to fill gaps in the acquired data. The assumptions that are 
made, typically that the cornea is spherical over any areas 
that cannot be imaged, are unwarranted (eg Klyce and 
Wilson 1989a, Bores 1991). 
e Photokeratoscopic data are highly sensitive to 
misalignment with the corneal axis (Heath et al 1991) and 
misjudgment of the focussing position and therefore the 
distance to the cornea (Saarloos and Constable 1991, 
Missotten 1994). 
e The instruments rely on corneal reflectivity, in turn 
determined by the condition of the unstable tear film that 
coats the surface of a healthy cornea and the surface 
roughness of the corneal epithelium (Duke-Elder 1970). 
For a healthy cornea, reflectivity is only about 4% at the 
corneal centre and decreases to near 2% at the periphery 
(Clark 1973a). Abnormal corneas with low reflectivity 
cannot be measured (Warnicki et al 1988). 
Photokeratoscopes cannot be used for assessment of 
corneal topography during surgery because of the 
inevitable non-reflectivity of the corneal surface. 
e Exact topographic data cannot be calculated from a 
photokeratoscope image. The curve fitting techniques 
applied are of limited value for several reasons but 
primarily because of the non-uniqueness of the corneal 
surface for a given image (eg Wise et al 1986, Mammone 
et al 1990), the asphericity of the corneal surface, and 
because they cannot model abrupt changes that may occur, 
for example, at the edge of a photorefractive keratectomy 
(Missotten 1994). 
3. THE KERATOCON 
The limited quantitative information provided by the 
keratometer and photokeratoscope together with an 
increasing requirement in modern ophthalmology for 
accurate topographic mapping of the entire anterior surface 
of both normal and abnormal corneas made it apparent that a 
new instrument was required. It follows from the discussions 
above that such an instrument should: 
i. be capable of measuring the entire cornea, 
ii. not rely on corneal reflectance, 
iii. not rely on a precorneal tear film, 
iv. measure corneal topography with sufficient density and 
accuracy to provide reliable and clinically interpretable 
representations of corneal topography and corneal 
power, and 
v. not require that assumptions be made about the 
geometry of the cornea. 
A schematic of the instrument appears below. The cameras 
used were two 35mm motor-driven Leica R4's fitted with 
Leica 200mm focal length lenses, positioned at 
approximately 25° convergence. Because the cameras were 
non-metric it was necessary to design the prototype so that 
on-the-job calibrations could be performed. This required 
photocontrol very close to the cornea — within the limited 
depth of field of both cameras and sufficiently close to be 
photogrammetrically reliable. There are two operational 
considerations. Firstly, the clinician must have access to the 
eye; secondly, patients cannot be expected to tolerate close 
proximity to any part of the instrument. To overcome these 
problems, photocontrol was reflected from a beamsplitter 
into the optical axis of the system, so that it appeared to 
surround the cornea in each photograph. A similar technique 
has been used by Scott (1981, 1987) for his reflex measuring 
instruments and reported by Mikhail (1974). The control 
points were a pattern of 30 marks burnt through a thin 
opaque surface deposited onto the outside of an accurately 
polished glass sphere. The sphere is flash illuminated from 
behind at the moment that the cameras are fired. The control 
points were coordinated to better than +5um. The cameras, 
photocontrol and beamsplitter were mounted onto the 
platform of a Sun PKS1000 photokeratoscope. This 
provided a mechanism for controlling the height of the 
instrument and the position of the patient's head relative to 
the cameras. 
  
  
          
beamsplitter 
ri 
"4A ES | 
FI RSS A 
0 IRR 
POL KXX] 
bod 12292 li 
  
  
  
  
7777 gon 
pivot 
  
  
  
  
  
  
  
Figure 1. Schematic of the Keratocon. 
International Archives of Photogrammetry and Remote Sensing. Vol. XXXI, Part B5. Vienna 1996 
  
A variety 
including 
1961, B: 
particula) 
as the : 
Warnicki 
method a 
contact 1 
prototype 
1993 per 
indicatioi 
simpler t 
onto it. 
redirect 
where m 
cornea to 
bundle fi 
control sj 
  
Figur 
in the 
The calib 
using bot 
and Kara 
Karara 1¢ 
routine d. 
space con 
monosco] 
prior to p: 
It was in 
accuracy 
effects of 
base distz 
of object 
constraine 
of field cc 
patient's | 
found to 
and conve 
Formulae 
Abdel-Az 
precision 
geometrie 
l  obje 
the ii
	        