time the cornea is photographed.
slitlamp was used to visually check the adherence of the
lens to the corneal surface for every patient.
Comparisons of corneas measured repeatedly and
comparisons of Keratocon and photokeratoscope
measurements also indicate that the surface of the
topography is being accurately represented by the
contact lens. The most difficult aspect to measure is the
sensitivity of the solution to rapid changes in corneal
topography.
The cornea is mapped in a coordinate system defined by
the object space control. This means that, although an
X-Y-Z translation of the patients head between
successive measurements introduces no errors, it is
important that the patient fixates the same point each
Figure 4. Photokeratoscope image of a patient with
conical cornea. Irregular mires can be seen and an
impression of downward displacement of the cone. This
cornea was too steep to measure with a keratometer.
0 5mm
| 1 |
0.1mm Contour Interval
Figure 5. Corneal map showing the displacement of the
cone and the flattening of the cornea.
The Keratocon was installed at the Tasmanian Lion's Eye
Diagnostic Centre and tested on a selection of patients with
corneal abnormalities. Figures 4 and 5 are of a patient with a
conical left cornea. The contours show the disturbed
topography, known as a "droopy cone". The cornea is
flattened above the apex and steeper inferiorly, with temporal
displacement of the apex. The photokeratoscope provides
very little information about this patient. The cornea was too
steep to measure using keratometry. The Keratocon has
mapped the cornea to its periphery whereas the
photokeratoscope has imaged less than 50% of the corneal
surface.
4. A DIGITAL KERATOCON
To meet with clinical acceptance, a biomedical measurement
system must satisfy the following conditions:
i. it must not discomfort or intimidate the patient,
ii. it must only require the patient's cooperation for a short
period of time,
iii. it must be reliable,
iv. it must output measurement parameters that suit the
clinician,
v. it must present those parameters in a readily
understandable form,
vi. it must be simple to use,
vii. it must be accurate, and
viii. the time taken to process and present data must suit the
application.
The relative importance of each of these conditions will
depend upon the application but in broad terms, and certainly
in the case of any method of measuring corneal topography,
the order in which these have been listed is indicative of their
importance.
Accuracy may be the least important of the issues. In the
case of corneal measurement this is illustrated by the success
of keratoscopic methods. Speed, except in the case of
intraoperative procedures, is also not absolutely necessary.
The turn around time on clinical procedures such as
angiograms and X-rays indicate the delay that clinicians are
prepared to tolerate. Ease of use is important, but
examination of existing ophthalmic instrumentation such as
keratoscopes and slit lamps suggests that clinicians will
tolerate quite poor ease of use. The most critical
determinants of an instrument's success in a normal clinical
environment appear to be the first four on the list, namely the
reliable and easily interpreted presentation of appropriate
information from an instrument that minimises patient
discomfort. Appropriate information includes estimates of
how the measured corneal abnormalities will affect vision
and the information needed in order that consequent vision
defects can be corrected.
A digital prototype of the keratocon should therefore meet
the following criteria:
i. sufficient metric reliability to avoid the need for on-the-
job calibrations,
ii. real time video imagery for patient alignment,
iii. automatic or semi-automatic target recognition and
measurement,
iv. automatic computation of three dimensional data,
v. high data reliability,
International Archives of Photogrammetry and Remote Sensing. Vol. XXXI, Part B5. Vienna 1996
vi. hig
inf
vii. ne
ap|
The inv
is a cri
typical
reduced
to signi
compen
flow f
relative
limit su.
digital «
would |
which n
15mm 3
betweer
framegr
from a
techniqi
photogr
allow tl
time pri
clinical
therefor
image.
In the c
with 10
scale 1:
diamete
accurac
than +4
analogu
expecte
Fig
COIT
Target ;
aspect o
has beei
of autor
targeted
Scanned
The im
histogra
is reasc