Full text: Resource and environmental monitoring

  
  
  
sum value has been chosen. 
Pulse 7 
  
  
Arbitrary Units 
& 
  
  
  
Sample 
  
  
  
Figure 4: Pulse 7 Peak Enlargement 
The values of the distances obtained by the DIAL system are 
equal, respectively, to 339.3 and 1086.0 m; the relevant standard 
deviations are equal to 0.73 and 1.47 m. The values obtained 
with the total station are equal to 337.579 and 1085.231 m. 
The DIAL facility is generally used for the RS of atmospheric 
parameters, so, a test with gaseous targets has been made. 
The concentration of water vapour has been obtained with the 
DIAL system, along a straight line, connecting the system mirror 
and a building of the university. The laser beams crossed the 
atmosphere on the botanic garden, delimited by a road and a 
built zone (Figure 5). 
  
Figure 5: Layout of the University of Calabria 
Some variations of concentration have been detected, in 
corrispondence of the buildings and of the botanic garden; the 
DIAL signals are shown in the Figure 6. Figure 7 shows the H,O 
concentration variations along the laser beams path. 
  
  
ON 
un a a ORV ^" 
  
  
  
  
  
  
  
  
  
a 
= 
2 
» 
£ 
onpazess23883822 
- AA N oO FT © 
Range 
  
  
—, 
Figure 6: Dial signals used for Water Vapour measurement 
  
  
  
  
  
  
  
Water Vapour 
20000 
15000 1 
5 
2 10000 + 
5000 | 
0 4 
© 0 o Le) Ww wn o 
E. Nut. aS alis 
+ e) wn 
Range 
  
Figure 7: HyO concentration 
The difference between the ranges obtained by DIAL system and 
total station is about 10 meters. In this case, to obtain the water 
vapour concentration, it is necessary to consider a 20 m cell, to 
have a good evaluation of the gradients of the lj, and loft 
signals, as shown in the DIAL equation. 
Angle measurements 
The first problem to solve is the alignment of the laser beams 
with the telescope axis. For this aim, a HeNe laser beam, 
previously aligned with the telescope axis, has been projected at 
a distance of about 15 m and a red spot has been obtained on a 
wall of the laboratory. In this point, an IR target (thermal 
printing paper) has been positioned. The energy of the CO» laser 
pulses is sufficient to generate a black spot on the paper. Two 
mirrors, allowing micrometric movements, have been tilted to 
superimpose the black spot on the red one. In this way, the 
6 Intemational Archives of Photogrammetry and Remote Sensing. Vol. XXXII, Part 7, Budapest, 1998 
alignmen 
about 0.1 
The seco 
measuren 
0.4 m, | 
(Cassegr: 
it is diffi 
circles al 
beams. 
eee 
  
The resul 
HeNe la: 
one sexa 
To obtai 
beams, a 
It will be 
the labor: 
Two mot 
vertical | 
horizonta 
collimati 
In the n 
vapour ti 
plan).
	        